Sign up for our newsletter

My Cart  

My Account

Language

Currency

Rhodamine 6G

Rhodamine 6G

Rhodamine is commonly used as a dye for tagging purposes. 

Rhodamine 6G has three important corresponding peaks in Raman spectroscopy, namely: 1310, 1360, 1507 cm-1

Fig 1. info

Information:
Laser wavelength: 785 nm

Laser power: 0.4 mW

Integration time: 10 seconds

Lens magnification: 50x

Raman spectrometer: Renishaw

Concentration: 0.1 mM of Rhodamine 6G (dried)

Substrate used:  Silver

 

Reproducibility

Information:

Laser wavelength: 532 nm

Laser power: 0.16 mW

Integration time: 1 second

Lens magnification: 50x

Raman spectrometer: Renishaw

Substrate used:  Silver

Concentration: 0.1 mM of Rhodamine 6G (dried)

 

 

 

Information:

Laser wavelength: 785 nm

Laser power: 0.4 mW

Integration time: 10 seconds

Lens magnification: 50x

Raman spectrometer: Renishaw

Concentration: 0.1 mM of Rhodamine 6G (wet)

Substrate used:  Silver

 

Information:

Laser wavelength: 785 nm

Laser power: 0.4 mW

Integration time: 1 second

Lens magnification: 50x

Raman spectrometer: Renishaw

Concentration: 0.1 mM of Rhodamine 6G (dried)

Substrate used:  Silver